Z-0683-2020 Class II Terminated
FDA device recall Z-0683-2020 was initiated by Polymer Technology Systems on Dec 2, 2016 and is designated Class II. Reason for recall: A numerical value less than 4% or greater than 13% may be displayed instead of the expected <4% or >13% result, due to a software bug. The recall status is terminated (terminated Jul 16, 2020). Affected quantity: 500.
Recall details
- Product Type
- Devices
- Report Date
- Dec 25, 2019
- Initiation Date
- Dec 2, 2016
- Termination Date
- Jul 16, 2020
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 500
Product description
Reason for recall
Distribution pattern
The products were distributed to the following US states: AR, CA, FL, IL, IN, KY, MN, NC, NJ, NY, OH, PA, RI TN, UT, VA, WA, and Puerto Rico. The products were distributed to the following foreign countries: Australia, Bahamas, Belgium, Bahrain, Bangladesh, Canada, China, Columbia, Czech Republic, England, Egypt, Greece, Spain, India, Indonesia, Haiti, Hong Kong, Italy, Japan, Jordan, Kenya, Lebanon, Mexico, Myanmar, Malaysia, Philippines, Pakistan, South Africa, Nigeria, Poland, Saudi Arabia, Sweden, Turkey, Singapore, Trinidad & Tobago, Slovenia, Vietnam, United Arab Emirates, Uruguay.
Code information
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-0683-2020
- openFDA file date
- Nov 19, 2026
- Loaded by RegDataLab
- Nov 20, 2026
Need the full FDA record for Polymer Technology Systems? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.