Z-0685-2017 Class III Terminated

Recalled by MicroPort Orthopedics Inc. — Arlington, TN

FDA device recall Z-0685-2017 was initiated by MicroPort Orthopedics Inc. on Jun 21, 2016 and is designated Class III. Reason for recall: Revision rate trends from within the MicroPort Orthopedics Inc. database evaluated for the modular head Metal on Metal THA systems show an increasing overall trend from 2009 to present and it was… The recall status is terminated (terminated Jul 1, 2017). Affected quantity: 500 units.

Recall details

Product Type
Devices
Report Date
Dec 7, 2016
Initiation Date
Jun 21, 2016
Termination Date
Jul 1, 2017
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
500 units

Product description

Procotyl (Rim- Lock) Liners: REF PHA04702, REF PHA04704, REF PHA04706, REF PHA04708, REF PHA04710, REF PHA04712, REF PHA04714 For use with the conserve metal head

Reason for recall

Revision rate trends from within the MicroPort Orthopedics Inc. database evaluated for the modular head Metal on Metal THA systems show an increasing overall trend from 2009 to present and it was found that there was a specific hazard/harm for suspected tissue reaction to metal debris over one percent for modular head CONSERVE(R) systems and DYNASTY(R) metal acetabular liner systems.

Distribution pattern

Worldwide Distribution - US (nationwide) Argentina, Australia, Austria, Belgium, Brazil, Bulgaria, Canada, China, Columbia, Cyprus, Czech Republic, Denmark, Finland, France, Georgia, Germany, Greece, Iran, Italy, Jamaica, Japan, Lithuania, Malaysia, Philippines, Poland, Portugal, Russia, Slovenia, South Africa, South Korea, Spain, Switzerland, Taiwan, Netherlands, Turkey, Arab Emirates, United Kingdom, and Vietnam.

Code information

REF PHA04702 Serial Numbers: PHA04702, 087461467, 089909592, 106368684, 106379196, 106381827, 109949208, 129989153, 0201018636, 0301049040 REF PHA04704 Serial Numbers: 010989155, 027373842, 038461468, 057433172, 077453350, 089924904, 106379197, 106380366, 106381751, 0201018637 REF PHA04706 Serial Numbers: 1414640, 010990534, 017400915, 037415861, 039755749, 047420835, 047424536, 077459275, 087461469, 089906466, 0101034747, 106379879, 106381773, 0301084459, 0401100079, 0501122040, 0801189074 REF PHA04708 Serial Numbers: 010990535, 017400568, 037420836, 047424415, 077455537, 087461470, 089901226, 106380367, 106381752, 0311304891, 0401118999, 0601163086, 0711360334 REF PHA04710 Serial Numbers: 1372615, 1413361, 017400673, 037415862, 037420837, 047423789, 047424535, 077457450, 077460608, 089901146, 098695744, 098695746, 106380368, 106381542, 106381753, 108695745, 109944906, 116380977, 119983782, 129990536, 0201016259, 0301076759, 0311303536, 0311305118, 0401089452, 0401121936, 0501131023, 0801187747, 1291000322 REF PHA04712 Serial Numbers: 029695877, 037414678, 047418090, 057426407, 077457328, 098464129, 099906467, 106373846, 106380571, 116389410, 0201058381, 0301084465, 0311303537, 0401117709, 0601163076, 0801188639 REF PHA04714 Serial Numbers: 1414641, 017400569, 029807399, 037415854, 047423790, 077432883, 077445957, 106368690, 0201025145, 0901196483

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0685-2017
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for MicroPort Orthopedics Inc.? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-0685-2017. Report a data issue: hello@regdatalab.com.