Z-0685-2017 Class III Terminated
Recalled by MicroPort Orthopedics Inc.
— Arlington, TN
FDA device recall Z-0685-2017 was initiated by MicroPort Orthopedics Inc. on Jun 21, 2016 and is designated Class III. Reason for recall: Revision rate trends from within the MicroPort Orthopedics Inc. database evaluated for the modular head Metal on Metal THA systems show an increasing overall trend from 2009 to present and it was… The recall status is terminated (terminated Jul 1, 2017). Affected quantity: 500 units.
Recall details
- Product Type
- Devices
- Report Date
- Dec 7, 2016
- Initiation Date
- Jun 21, 2016
- Termination Date
- Jul 1, 2017
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 500 units
Product description
Procotyl (Rim- Lock) Liners: REF PHA04702, REF PHA04704, REF PHA04706, REF PHA04708, REF PHA04710, REF PHA04712, REF PHA04714 For use with the conserve metal head
Reason for recall
Revision rate trends from within the MicroPort Orthopedics Inc. database evaluated for the modular head Metal on Metal THA systems show an increasing overall trend from 2009 to present and it was found that there was a specific hazard/harm for suspected tissue reaction to metal debris over one percent for modular head CONSERVE(R) systems and DYNASTY(R) metal acetabular liner systems.
Distribution pattern
Worldwide Distribution - US (nationwide) Argentina, Australia, Austria, Belgium, Brazil, Bulgaria, Canada, China, Columbia, Cyprus, Czech Republic, Denmark, Finland, France, Georgia, Germany, Greece, Iran, Italy, Jamaica, Japan, Lithuania, Malaysia, Philippines, Poland, Portugal, Russia, Slovenia, South Africa, South Korea, Spain, Switzerland, Taiwan, Netherlands, Turkey, Arab Emirates, United Kingdom, and Vietnam.
Code information
REF PHA04702 Serial Numbers: PHA04702, 087461467, 089909592, 106368684, 106379196, 106381827, 109949208, 129989153, 0201018636, 0301049040 REF PHA04704 Serial Numbers: 010989155, 027373842, 038461468, 057433172, 077453350, 089924904, 106379197, 106380366, 106381751, 0201018637 REF PHA04706 Serial Numbers: 1414640, 010990534, 017400915, 037415861, 039755749, 047420835, 047424536, 077459275, 087461469, 089906466, 0101034747, 106379879, 106381773, 0301084459, 0401100079, 0501122040, 0801189074 REF PHA04708 Serial Numbers: 010990535, 017400568, 037420836, 047424415, 077455537, 087461470, 089901226, 106380367, 106381752, 0311304891, 0401118999, 0601163086, 0711360334 REF PHA04710 Serial Numbers: 1372615, 1413361, 017400673, 037415862, 037420837, 047423789, 047424535, 077457450, 077460608, 089901146, 098695744, 098695746, 106380368, 106381542, 106381753, 108695745, 109944906, 116380977, 119983782, 129990536, 0201016259, 0301076759, 0311303536, 0311305118, 0401089452, 0401121936, 0501131023, 0801187747, 1291000322 REF PHA04712 Serial Numbers: 029695877, 037414678, 047418090, 057426407, 077457328, 098464129, 099906467, 106373846, 106380571, 116389410, 0201058381, 0301084465, 0311303537, 0401117709, 0601163076, 0801188639 REF PHA04714 Serial Numbers: 1414641, 017400569, 029807399, 037415854, 047423790, 077432883, 077445957, 106368690, 0201025145, 0901196483
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-0685-2017
- openFDA file date
- Nov 25, 2026
- Loaded by RegDataLab
- Nov 26, 2026
These dates say when openFDA published the data file and when RegDataLab last loaded it.
Need the full FDA record for MicroPort Orthopedics Inc.? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.
Get a free API key
How the connector works