Z-0686-2017 Class III Terminated

Recalled by MicroPort Orthopedics Inc. — Arlington, TN

FDA device recall Z-0686-2017 was initiated by MicroPort Orthopedics Inc. on Jun 21, 2016 and is designated Class III. Reason for recall: Revision rate trends from within the MicroPort Orthopedics Inc. database evaluated for the modular head Metal on Metal THA systems show an increasing overall trend from 2009 to present and it was… The recall status is terminated (terminated Jul 1, 2017). Affected quantity: 22 units.

Recall details

Product Type
Devices
Report Date
Dec 7, 2016
Initiation Date
Jun 21, 2016
Termination Date
Jul 1, 2017
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
22 units

Product description

Lineage Shell: REF 36693646, REF 36693848, REF 36694050, REF 36694252, REF 36694454, REF 36694656, REF 36694858, REF 36695060, REF 36695262, REF 36695464 For use with the conserve metal head

Reason for recall

Revision rate trends from within the MicroPort Orthopedics Inc. database evaluated for the modular head Metal on Metal THA systems show an increasing overall trend from 2009 to present and it was found that there was a specific hazard/harm for suspected tissue reaction to metal debris over one percent for modular head CONSERVE(R) systems and DYNASTY(R) metal acetabular liner systems.

Distribution pattern

Worldwide Distribution - US (nationwide) Argentina, Australia, Austria, Belgium, Brazil, Bulgaria, Canada, China, Columbia, Cyprus, Czech Republic, Denmark, Finland, France, Georgia, Germany, Greece, Iran, Italy, Jamaica, Japan, Lithuania, Malaysia, Philippines, Poland, Portugal, Russia, Slovenia, South Africa, South Korea, Spain, Switzerland, Taiwan, Netherlands, Turkey, Arab Emirates, United Kingdom, and Vietnam.

Code information

REF 36693646 Serial Numbers: 09230967, 09229086, 10230508 REF 36693848 Serial Numbers: 08228745, 10230509, 11235075 REF 36694050 Serial Numbers: 09229091, 09230501 REF 36694252 Serial Numbers: 09229131, 10230510 REF 36694454 Serial Numbers: LLALMB161A, 09230503 REF 36694656 Serial Numbers: 09230505, 11230847 REF 36694858 Serial Numbers: 09229167, 10230511 REF 36695060 Serial Numbers: 09229169, 09230506 REF 36695262 Serial Numbers: 09230507, 09229170 REF 36695464 Serial Numbers: 08228747, 10230512

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0686-2017
openFDA file date
Nov 16, 2026
Loaded by RegDataLab
Nov 17, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-0686-2017. Report a data issue: hello@regdatalab.com.