Z-0701-2016 Class II Terminated

Recalled by Mindray DS USA, Inc. Dba Mindray North America — Mahwah, NJ

FDA device recall Z-0701-2016 was initiated by Mindray DS USA, Inc. Dba Mindray North America on Dec 15, 2015 and is designated Class II. Reason for recall: An issue with the Panorama Central Station may cause the system to spontaneously restart. This may occur after about 49 days of continuous operation. On restart, monitoring will be lost for about 2.5… The recall status is terminated (terminated Mar 29, 2017). Affected quantity: 63 products.

Recall details

Product Type
Devices
Report Date
Feb 10, 2016
Initiation Date
Dec 15, 2015
Termination Date
Mar 29, 2017
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
63 products

Product description

Panorama Patient Monitoring Network. Software Kit 0020-00-0205-19A, View Station Software Kit 0020-00-0206-16A, Work Station Software Kit 0020-00-0207-16A, Software Kit Upgrade TECH-00-910. The Panaroma Patient Monitoring Network is intended for use in a fixed location in the healthcare facility setting as a central viewing station. It is not intended to be directly connected to patient at any time or installed in a patient's vicinity.

Reason for recall

An issue with the Panorama Central Station may cause the system to spontaneously restart. This may occur after about 49 days of continuous operation. On restart, monitoring will be lost for about 2.5 minutes. Settings and configurations are maintained.

Distribution pattern

Nationwide Distribution.

Code information

version 11.5.1

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0701-2016
openFDA file date
Nov 10, 2026
Loaded by RegDataLab
Nov 11, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-0701-2016. Report a data issue: hello@regdatalab.com.