Z-0713-2024 Class II Ongoing
FDA device recall Z-0713-2024 was initiated by mo-Vis BVBA on November 28, 2023 and is designated Class II. Reason for recall: In rare cases, the lever of the throttle can get temporarily stuck during the movement and does not come back to its neutral point by itself. If this happens during usage, an unwanted movement of the … The recall status is ongoing. Affected quantity: 129 units.
Recall Details
- Product Type
- Devices
- Report Date
- January 17, 2024
- Initiation Date
- November 28, 2023
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 129 units
Product Description
Scoot Control R-net, Product Number P015-61. The Scoot Control is intended to support a wheelchair attendant in controlling or maneuvering a power chair
Reason for Recall
In rare cases, the lever of the throttle can get temporarily stuck during the movement and does not come back to its neutral point by itself. If this happens during usage, an unwanted movement of the chair could be the result. An unwanted movement could result in damage to the wheelchair, material damage around the chair and/or injury to the user, attendant or by-standers.
Distribution Pattern
US States: TN, TX.
Code Information
UDI-DI: 05407008320119; Lot Numbers: 3304-3317 3322-3329 3440-3452 3569-3618 3830-3844 4020-4054 4170-4176 4250-4255 4321-4330