Z-2459-2025 Class I Ongoing

Recalled by mo-Vis BVBA — Nevele, N/A

Recall Details

Product Type
Devices
Report Date
September 10, 2025
Initiation Date
July 10, 2025
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
25 units

Product Description

IDM-ARLITE-R. Electrical wheelchair component.

Reason for Recall

Devices with firmware versions 2.3 and lower may experience timing variations between software tasks, which could result in skipping the necessary neutral check and the wheelchair producing unintended movement upon power-up when the joystick is out of neutral.

Distribution Pattern

US Nationwide distribution in the state of TX.

Code Information

Model No. PRSPS00018. UDI-DI: 05407008320102; Serial No. 1000 to 2647.