Z-2456-2025 Class I Ongoing
FDA device recall Z-2456-2025 was initiated by Mo-Vis BVBA on Jul 10, 2025 and is designated Class I. Reason for recall: Devices with firmware versions 2.3 and lower may experience timing variations between software tasks, which could result in skipping the necessary neutral check and the wheelchair producing unintended… The recall status is ongoing. Affected quantity: 15 units.
Recall details
- Product Type
- Devices
- Report Date
- Sep 10, 2025
- Initiation Date
- Jul 10, 2025
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 15 units
Product description
Reason for recall
Distribution pattern
US Nationwide distribution in the state of TX.