Z-2456-2025 Class I Ongoing

Recalled by Mo-Vis BVBA — Nevele, N/A

FDA device recall Z-2456-2025 was initiated by Mo-Vis BVBA on Jul 10, 2025 and is designated Class I. Reason for recall: Devices with firmware versions 2.3 and lower may experience timing variations between software tasks, which could result in skipping the necessary neutral check and the wheelchair producing unintended… The recall status is ongoing. Affected quantity: 15 units.

Recall details

Product Type
Devices
Report Date
Sep 10, 2025
Initiation Date
Jul 10, 2025
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
15 units

Product description

IDM-MICRO-R. Electrical wheelchair component.

Reason for recall

Devices with firmware versions 2.3 and lower may experience timing variations between software tasks, which could result in skipping the necessary neutral check and the wheelchair producing unintended movement upon power-up when the joystick is out of neutral.

Distribution pattern

US Nationwide distribution in the state of TX.

Code information

Model No. PRSPS0014; UDI-DI: 05407008320836; Serial No. 1000 to 1690.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2456-2025
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Mo-Vis BVBA? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Mo-Vis BVBA: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Mo-Vis BVBA →

Source: openFDA record for Z-2456-2025. Report a data issue: hello@regdatalab.com.