Z-2456-2025 Class I Ongoing
Recall Details
- Product Type
- Devices
- Report Date
- September 10, 2025
- Initiation Date
- July 10, 2025
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 15 units
Product Description
IDM-MICRO-R. Electrical wheelchair component.
Reason for Recall
Devices with firmware versions 2.3 and lower may experience timing variations between software tasks, which could result in skipping the necessary neutral check and the wheelchair producing unintended movement upon power-up when the joystick is out of neutral.
Distribution Pattern
US Nationwide distribution in the state of TX.
Code Information
Model No. PRSPS0014; UDI-DI: 05407008320836; Serial No. 1000 to 1690.