Z-0737-2020 Class II Terminated

Recalled by Diagnostica Stago, Inc. — Parsippany, NJ

FDA device recall Z-0737-2020 was initiated by Diagnostica Stago, Inc. on Nov 18, 2019 and is designated Class II. Reason for recall: A positive bias was confirmed to exists on the STA - Unicalibrator levels with STA - Deficient IX and STA - Deficient VIII when compared to the International Standard (IS) for factor VIII or IX. This… The recall status is terminated (terminated Jan 12, 2023). Affected quantity: 5,498.

Recall details

Product Type
Devices
Report Date
Jan 1, 2020
Initiation Date
Nov 18, 2019
Termination Date
Jan 12, 2023
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
5,498

Product description

STA UNICALIBRATOR (ref. 00675)

Reason for recall

A positive bias was confirmed to exists on the STA - Unicalibrator levels with STA - Deficient IX and STA - Deficient VIII when compared to the International Standard (IS) for factor VIII or IX. This bias is also observed on several External Quality Assessment (EQA) results.

Distribution pattern

US: MA ME VT CT NJ NY PA DC MD VA NC SC GA FL AL TN MS KY OH IN MI IA WI MN SD ND MT IL MO KS LA AR OK TX CO AZ NV CA HI OR WA AK

Code information

REF: 00675 Lot Numbers: 253180, 253964, 254721, 255108 UDIs: (01)13607450006756(11)171231(17)191231(10)253180(241)00675 (01)13607450006756(11)180630(17)200630(10)253964(241)00675 (01)13607450006756(11)181130(17)201130(10)254721(241)00675 (01)13607450006756(11)190228(17)210228(10)255108(241)00675

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0737-2020
openFDA file date
Nov 16, 2026
Loaded by RegDataLab
Nov 17, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-0737-2020. Report a data issue: hello@regdatalab.com.