Z-0737-2020 Class II Terminated
FDA device recall Z-0737-2020 was initiated by Diagnostica Stago, Inc. on Nov 18, 2019 and is designated Class II. Reason for recall: A positive bias was confirmed to exists on the STA - Unicalibrator levels with STA - Deficient IX and STA - Deficient VIII when compared to the International Standard (IS) for factor VIII or IX. This… The recall status is terminated (terminated Jan 12, 2023). Affected quantity: 5,498.
Recall details
- Product Type
- Devices
- Report Date
- Jan 1, 2020
- Initiation Date
- Nov 18, 2019
- Termination Date
- Jan 12, 2023
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 5,498
Product description
STA UNICALIBRATOR (ref. 00675)
Reason for recall
A positive bias was confirmed to exists on the STA - Unicalibrator levels with STA - Deficient IX and STA - Deficient VIII when compared to the International Standard (IS) for factor VIII or IX. This bias is also observed on several External Quality Assessment (EQA) results.
Distribution pattern
US: MA ME VT CT NJ NY PA DC MD VA NC SC GA FL AL TN MS KY OH IN MI IA WI MN SD ND MT IL MO KS LA AR OK TX CO AZ NV CA HI OR WA AK
Code information
REF: 00675 Lot Numbers: 253180, 253964, 254721, 255108 UDIs: (01)13607450006756(11)171231(17)191231(10)253180(241)00675 (01)13607450006756(11)180630(17)200630(10)253964(241)00675 (01)13607450006756(11)181130(17)201130(10)254721(241)00675 (01)13607450006756(11)190228(17)210228(10)255108(241)00675
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-0737-2020
- openFDA file date
- Nov 16, 2026
- Loaded by RegDataLab
- Nov 17, 2026