Z-0738-2020 Class II Terminated

Recalled by Diagnostica Stago, Inc. — Parsippany, NJ

Recall Details

Product Type
Devices
Report Date
January 1, 2020
Initiation Date
November 18, 2019
Termination Date
January 12, 2023
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
5,437

Product Description

STA SYSTEM CONROL N + P (ref. 00678)

Reason for Recall

A positive bias was confirmed to exists on the STA - Unicalibrator levels with STA - Deficient IX and STA - Deficient VIII when compared to the International Standard (IS) for factor VIII or IX. This bias is also observed on several External Quality Assessment (EQA) results.

Distribution Pattern

US: MA ME VT CT NJ NY PA DC MD VA NC SC GA FL AL TN MS KY OH IN MI IA WI MN SD ND MT IL MO KS LA AR OK TX CO AZ NV CA HI OR WA AK

Code Information

REF: 00678 Lot Numbers: 253145, 253416, 253655, 254040, 254206, 254401, 254573, 254760 UDIs: (01)03607450006780(11)171130(17)191130(10)253145(241)00678 (01)03607450006780(11)180131(17)200131(10)253416(241)00678 (01)03607450006780(11)180331(17)200331(10)253655(241)00678 (01)03607450006780(11)180630(17)200630(10)254040(241)00678 (01)03607450006780(11)180731(17)200731(10)254206(241)00678 (01)03607450006780(11)180930(17)200930(10)254401(241)00678 (01)03607450006780(11)181031(17)201031(10)254573(241)00678 (01)03607450006780(11)181130(17)201130(10)254760(241)00678