Z-0740-2020 Class II Terminated

Recalled by Ortho Clinical Diagnostics Inc — Raritan, NJ

Recall Details

Product Type
Devices
Report Date
January 1, 2020
Initiation Date
October 14, 2019
Termination Date
November 13, 2020
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
2,573

Product Description

Urine Electrolyte Diluent, Product Code: 1112352 - Product Usage: used to dilute urine specimens for analysis of sodium (Na+) and potassium (K+) on VITROS 250/350/5,1 FS/4600 Chemistry Systems and the VITROS 5600/XT 7600 Integrated Systems.

Reason for Recall

Potential for the instability of the diluents VITROS 7% BSA (Product Code 8262487) and VITROS Urine Electrolyte Diluent (UED) (Product Code 1110352) when stored in a 2mL cup for 7 days.

Distribution Pattern

Worldwide distribution - US Nationwide distribution and countries BM, AU, BR, CA, CL, CN, CO, IN, JP, MX, SG, GB, FR, DE, IT, ES, PT, PL, RU, DK, NO, SE, BE, NL.

Code Information

Lot with Expiry Date: V6391 2020-02-26 W6895 2020-09-20 X7127 2021-01-17 Y7593 2021-08-28