Z-0803-2025 Class II Ongoing

Recalled by Beckman Coulter, Inc. — Chaska, MN

FDA device recall Z-0803-2025 was initiated by Beckman Coulter, Inc. on Nov 26, 2024 and is designated Class II. Reason for recall: Beckman Coulter has identified that some Access PTH reagent packs from lot 339071 may be over- or under-filled in well 0. The dispensed material in well 0 could be above or below specification… The recall status is ongoing. Affected quantity: 362 units.

Recall details

Product Type
Devices
Report Date
Jan 15, 2025
Initiation Date
Nov 26, 2024
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
362 units

Product description

Access Intact PTH assay, a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of intact parathyroid hormone (parathyrin, PTH) levels in human serum and plasma using the Access Immunoassay Systems, Catalog Number A16972

Reason for recall

Beckman Coulter has identified that some Access PTH reagent packs from lot 339071 may be over- or under-filled in well 0. The dispensed material in well 0 could be above or below specification, risking sealing issues, contamination, instrument errors, delays, or inaccurate results with a negative bias and increase in imprecision. The issue was confirmed on 10 April 2024 by Beckman Coulter through an internal non-conformance (NC-INT-66585). The well 0 scale on the filling line began malfunctioning on the fill immediately before PTH. The malfunction was not noticed until the majority of PTH was filled.

Distribution pattern

Worldwide distribution - US Nationwide and the country of Canada.

Code information

UDI: (01)15099590201937(17)241130(11)231201; Lot 339071

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0803-2025
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Beckman Coulter, Inc.? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Beckman Coulter, Inc.: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Beckman Coulter, Inc. →

Source: openFDA record for Z-0803-2025. Report a data issue: hello@regdatalab.com.