Z-0837-2015 Class II Terminated
FDA device recall Z-0837-2015 was initiated by Mindray DS USA, Inc. Dba Mindray North America on Oct 6, 2014 and is designated Class II. Reason for recall: Mindray has identified an issue that involves the IBP module used with the DPM 6 and DPM 7 Monitor where the IBP measurement data may be displayed in an incorrect IBP tile on the monitors display. The recall status is terminated (terminated Jun 6, 2017). Affected quantity: 455 units.
Recall details
- Product Type
- Devices
- Report Date
- Dec 31, 2014
- Initiation Date
- Oct 6, 2014
- Termination Date
- Jun 6, 2017
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 455 units
Product description
DPM 6 Patient Monitor and DPM 7 Patient Monitor. The product affected is the Mindray IBP module, P/N is 6800-30-50485 for use with Mindray DPM 6 and DPM 7 patient monitors . Intended to be used for monitoring, displaying, reviewing, storing and alarming of multiple physiological parameters.
Reason for recall
Mindray has identified an issue that involves the IBP module used with the DPM 6 and DPM 7 Monitor where the IBP measurement data may be displayed in an incorrect IBP tile on the monitors display.
Distribution pattern
Distributed to the states of AL, CA, CO, FL, GA, IA, IL, IN, LA. MA, MO, MS, MT, NC, ND, NY, OH, PA, SC, TX, VA, WA and WV.
Code information
DPM 6 - PN: C-047-000199-00 and DPM 7 - PN: C-047-000200-00, IBP Module P/N: 6800-30-50485
Source