Z-0960-2023 Class I Ongoing

Recalled by Getinge Usa Sales Inc — Wayne, NJ

FDA device recall Z-0960-2023 was initiated by Getinge Usa Sales Inc on Dec 22, 2022 and is designated Class I. Reason for recall: Due to a software bug, under certain conditions, pressure cannot be built up resulting in no ventilation. If gas delivery is stopped, a sustained decrease in delivered O2 concentration may lead to… The recall status is ongoing. Affected quantity: 8.016 Total (7.766 OUS; 250 US).

Recall Details

Product Type
Devices
Report Date
Feb 1, 2023
Initiation Date
Dec 22, 2022
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
8.016 Total (7.766 OUS; 250 US)

Product Description

Getinge Flow-c Anesthesia System, Part No. 6887700

Reason for Recall

Due to a software bug, under certain conditions, pressure cannot be built up resulting in no ventilation. If gas delivery is stopped, a sustained decrease in delivered O2 concentration may lead to hypoventilation and hypoxia.

Distribution Pattern

US nationwide and Worldwide Distribution: Domestic distribution to AL, AR, AZ, CA, CO, DC, FL, GA, IN, MA, MI, MO, NY, OH, OR, PA, TX, VA, WA, WV. Foreign distribution worldwide.

Code Information

Part # UDI 6887700 07325710009765 Serial Numbers 4003-5078 Software versions 4.8.0, 4.8.1, 4.8.2, and 4.8.3.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0960-2023
openFDA file date
Oct 6, 2026
Loaded by RegDataLab
Oct 12, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it. FDA may have changed the recall since.