Z-1043-2017 Class II Terminated

Recalled by The Binding Site Group, Ltd. — Birmingham

FDA device recall Z-1043-2017 was initiated by The Binding Site Group, Ltd. on Jul 5, 2016 and is designated Class II. Reason for recall: Batch numbers listed are reporting Lambda Free Light Chain results >114mg/L that are on average 33% lower than other Optilite Lambda Freelite batches. Therefore users moving to this batch will see… The recall status is terminated (terminated Jul 3, 2017). Affected quantity: 809 kits.

Recall details

Product Type
Devices
Report Date
Jan 25, 2017
Initiation Date
Jul 5, 2016
Termination Date
Jul 3, 2017
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
809 kits

Product description

Optilite Freelite Lambda Free Kit Product Code: LK018.OPT, LK018.OPT.A Protein testing kit.

Reason for recall

Batch numbers listed are reporting Lambda Free Light Chain results >114mg/L that are on average 33% lower than other Optilite Lambda Freelite batches. Therefore users moving to this batch will see results >114mg/L giving lower results and conversely when switching to the next batch results will be seen to shift upwards.

Distribution pattern

U.S. distribution to the following; AR, PA, GA, NC, KY, VA, CA. Foreign distribution to the following; Uk, Germany, France, Italy, Spain, Czech Republic, Portugal, Belgium.

Code information

Lot #405566, 405567

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1043-2017
openFDA file date
Nov 19, 2026
Loaded by RegDataLab
Nov 20, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for The Binding Site Group, Ltd.? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-1043-2017. Report a data issue: hello@regdatalab.com.