Z-1071-2017 Class II Terminated

Recalled by The Binding Site Group, Ltd. — Birmingham, N/A

FDA device recall Z-1071-2017 was initiated by The Binding Site Group, Ltd. on Sep 11, 2014 and is designated Class II. Reason for recall: Customer complaints identified that samples containing extremely high IgG4 levels exhibit antigen excess and will yield spuriously low IgG4 results. The recall status is terminated (terminated May 3, 2017).

Recall details

Product Type
Devices
Report Date
Feb 1, 2017
Initiation Date
Sep 11, 2014
Termination Date
May 3, 2017
Voluntary/Mandated
Voluntary: Firm initiated

Product description

Human lgG4 Subclass Liquid Reagent Kits for use on the Siemens BN"II Analyser Product Code: NK009.T This kit is intended for qualifying human IgG subclasses 1,2,3 and 4 immunoglobulins in serum on the Beckman Coulter AU series. Meaurement of these immunoglobulins ais in the diagnosis of abnormal protein metabolism and the body's lack of ability to resist infectious agents

Reason for recall

Customer complaints identified that samples containing extremely high IgG4 levels exhibit antigen excess and will yield spuriously low IgG4 results.

Distribution pattern

Worldwide Distribution - US including n FL, OH, TN, MN, NW, MI, CA. and Internationally to Canada, France, UK, Spain, Germany, Czech Republic, Belgium, Italy, Australia, Brazil, Chile, China, Costa Rica, Kuwait, New Zealand, Oman, Pakistan, Portugal, Sweded, Taiwan, and Turkey.

Code information

All lots.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1071-2017
openFDA file date
Nov 24, 2026
Loaded by RegDataLab
Nov 25, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for The Binding Site Group, Ltd.? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-1071-2017. Report a data issue: hello@regdatalab.com.