Z-1085-2022 Class II Ongoing

Recalled by Beckman Coulter Inc. — Brea, CA

FDA device recall Z-1085-2022 was initiated by Beckman Coulter Inc. on Oct 21, 2019 and is designated Class II. Reason for recall: WalkAway Instruments may have been assembled with the incorrect AC power entry connector inlet with is rated at 1 ampere current rating rather than the correct current rating of 10 ampere The recall status is ongoing. Affected quantity: 40 systems.

Recall details

Product Type
Devices
Report Date
May 25, 2022
Initiation Date
Oct 21, 2019
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
40 systems

Product description

DxM 1040 MicroScan WalkAway Instrument REF B1018-440

Reason for recall

WalkAway Instruments may have been assembled with the incorrect AC power entry connector inlet with is rated at 1 ampere current rating rather than the correct current rating of 10 ampere

Distribution pattern

Worldwide distribution - U.S. Nationwide distribution in the states of Alaska, Arkansas, California, Colorado, District of Columbia, Florida, Georgia, Idaho, Illinois, Indiana, Iowa, Kentucky, Louisiana, Maine, Maryland, Michigan, Minnesota, Missouri, New Jersey, New Mexico, New York, North Carolina, North Dakota, Ohio, Pennsylvania, Tennessee, Texas, Utah, Vermont, Virginia, and Washington. The countries of Belgium, Brazil, Colombia, Germany, Hong Kong, Indonesia, Italy, Japan, Paraguay, Philippines, Poland, Portugal, Romania, Russia, Saudi Arabia, Spain, Switzerland, Taiwan, United Arab Emirates, Vietnam and Yemen.

Code information

Catalog Number: B1018-440 UDI-DI Code: 15099590690885 Serial Numbers: 44010026 44010027 44010028 44010029 44010030 44010031 44010032 44010033 44010035 44010036 44010037 44010038 44010040 44010041 44010042 44010043 44010044 44010045 44010046 44010047 44010048 44010049 44010050 44010051 44010052 44010053 44010054 44010055 44010056 44010057 44010058 44010059 44010060 44010061 44010063 44010064 44010065 44010066 44010067 44010068

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1085-2022
openFDA file date
Nov 27, 2026
Loaded by RegDataLab
Nov 28, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.