Z-1120-2014 Class II Terminated

Recalled by Philips Medical Systems, Inc. — Andover, MA

FDA device recall Z-1120-2014 was initiated by Philips Medical Systems, Inc. on Feb 18, 2014 and is designated Class II. Reason for recall: Philips HeartStart MRx Monitor/Defibrillator, when operating on battery power only, may experience an unexpected shutdown if exposed to elevated levels of electromagnetic interference from RF energy… The recall status is terminated (terminated Sep 3, 2020). Affected quantity: 43,792 batteries.

Recall details

Product Type
Devices
Report Date
Mar 5, 2014
Initiation Date
Feb 18, 2014
Termination Date
Sep 3, 2020
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
43,792 batteries

Product description

HeartStart MRx Defibrillator/Monitor M3538A Lithium Ion Batteries Revision J The HeartStart MRx is for use for the termination of ventricular tachycardia and ventricular fibrillation. The device is for use by qualified medical personnel trained in the operation of the device and qualified by training in basic life support, advanced cardiac support, or defibrillation. It must be used by or on the order of a physician

Reason for recall

Philips HeartStart MRx Monitor/Defibrillator, when operating on battery power only, may experience an unexpected shutdown if exposed to elevated levels of electromagnetic interference from RF energy source

Distribution pattern

Worldwide Distribution: US (nationwide) and Internationally to: ALBANIA, ARGENTINA, ARUBA, AUSTRALIA, AUSTRIA, AZERBAIJAN, BAHRAIN, BANGLADESH, BELGUIM, BOLIVIA, BRAZIL, CANADA, CHINA, COLOMBIA, CONGO, CROATIA, CZECH REPUBLIC, DENMARK, EGYPT, EL SALVADOR, ESTONIA, ETHIOPIA, FINLAND, FRANCE, GERMANY, HONG KONG, ICELAND, INDIA, INDONESIA, IRELAND, ISRAEL, ITALY, JAPAN, JORDAN, KAZAKHSTAN, KENYA, KOREA REPUBLIC, KUWAIT, LATIVA, LEBANON, LIBYAN ARAB, LITHUANIA, MALAYSIA, MALTA, MEXICO, MONGOLIA, MOROCCO, MOZAMBIQUE, MYANMAR, NEPAL, NETHERLANDS, NEW CALEDONIA, NEW ZEALAND, NICARAGUA, NIGERIA, NORWAY, OMAN, PAKISTAN, PARAGUAY, PERU, PHILIPPINES, POLAND, PORTUGAL, QATAR, ROMANIA, RUSSIAN FEDERAL, SAUDIA ARABIA, SERBIA, SINGAPORE, SLOVENIA, SOUTH AFRICA, SPAIN, SWEDEN, SWITZERLAND, TAIWAN PROVIDENCE, THAILAND, TRINIDAND AND TABAGO, TUNISIA, TURKEY, TURKMENISTANIA, UKRAINE, UNITED ARAB EMIRATES, UNITED KINGDOM, UZBEKISTAN, VENEZUELA, and VIETNAM.

Code information

Serial numbers: 11327-0019-P 11315-0023-P 11315-0027-P 11315-0031-P 11315-0032-P 11315-0033-P 11315-0036-P 11315-0109-P 11315-0214-P 11315-0219-P 11315-0391-P 11322-0084-P 11322-0088-P 11327-0043-P 11327-0044-P Or within the following range: 11335-xxxx-P to 13214-xxxx-P

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1120-2014
openFDA file date
Nov 27, 2026
Loaded by RegDataLab
Nov 28, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Philips Medical Systems, Inc.? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Philips Medical Systems, Inc.: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Philips Medical Systems, Inc. →

Source: openFDA record for Z-1120-2014. Report a data issue: hello@regdatalab.com.