Z-1156-2014 Class II Terminated
FDA device recall Z-1156-2014 was initiated by Mindray DS USA, Inc. Dba Mindray North America on Jan 8, 2014 and is designated Class II. Reason for recall: Mindray has identified two issues with the Panorama Central Station. Full Disclosure data may lag behind the Panorama system time clock when the Panorama Central Station is in use with Mindray¿s… The recall status is terminated (terminated Jan 21, 2015). Affected quantity: 732 units in the US.
Recall details
- Product Type
- Devices
- Report Date
- Mar 19, 2014
- Initiation Date
- Jan 8, 2014
- Termination Date
- Jan 21, 2015
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 732 units in the US
Product description
Panaroma Central Station Product Usage: This device can view real time, store, print, graph, and trend patient clinical and demographic data. The device can also set independent alarm limits for data send by the bedside monitor.
Reason for recall
Mindray has identified two issues with the Panorama Central Station. Full Disclosure data may lag behind the Panorama system time clock when the Panorama Central Station is in use with Mindray¿s V-Series Monitor. Additionally, event indicators lag behind the Panorama system time clock when the Panorama Central Station is in use with a Panorama Telepack.
Distribution pattern
Worldwide Distribution - USA Nationwide Distribution
Code information
p/n 0998-00-0708-01; 0998-UC-0708-01 and 0020-00-0205-XXX