Z-1215-2015 Class II Terminated

Recalled by Brainlab AG — Feldkirchen, N/A

FDA device recall Z-1215-2015 was initiated by Brainlab AG on Jan 16, 2015 and is designated Class II. Reason for recall: The recommended sterilization and drying parameters are not effective to achieve the minimum required sterilization level. The recall status is terminated (terminated Mar 13, 2017). Affected quantity: 114 offset cup impactors.

Recall details

Product Type
Devices
Report Date
Mar 4, 2015
Initiation Date
Jan 16, 2015
Termination Date
Mar 13, 2017
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
114 offset cup impactors

Product description

The Brainlab Offset Cup Impactor Universal. Part of the Brainlab hip system. An intraoperative image-guided localization system to enable minimally invasive orthopedic surgery. (a stereotaxic instrument)

Reason for recall

The recommended sterilization and drying parameters are not effective to achieve the minimum required sterilization level.

Distribution pattern

Worldwide Distribution - US: Nationwide (AZ, CA, ID, IL, MD, MI, MN, OH, OK, OR, TN, WA, WI, WV); Australia, Austria, Bahrain, Belgium, Germany, Hong Kong, India, Italy, Japan, Malaysia, Russian Federation, Saudi Arabia, South Korea, Spain, Switzerland, Thailand, United Arab Emirates, United Kingdom

Code information

Model 52856 BL CURVED CUP INSERTER UNIVERSAL SHAFT; and MODEL 52856A BL OFFSET CUP IMPACTOR UNIVERSAL SHAFT, all serial/lot numbers distributed prior to January 13, 2015.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1215-2015
openFDA file date
Nov 19, 2026
Loaded by RegDataLab
Nov 20, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Brainlab AG? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-1215-2015. Report a data issue: hello@regdatalab.com.