Z-1249-2015 Class II Terminated
FDA device recall Z-1249-2015 was initiated by Microport Orthopedics INC. on Feb 2, 2015 and is designated Class II. Reason for recall: Some units of the EVOLUTION Tibial Base was affected by a casting tool issue which caused varying retaining wall thickness and keel offset in the finished product. The recall status is terminated (terminated Mar 19, 2015). Affected quantity: 17 units.
Recall details
- Product Type
- Devices
- Report Date
- Mar 18, 2015
- Initiation Date
- Feb 2, 2015
- Termination Date
- Mar 19, 2015
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 17 units
Product description
Reason for recall
Distribution pattern
Worldwide Distribution - USA including AL, AZ, ID, KS, KS, MO, TX, IA, WV and Internationally to Canada and The Netherlands.
Code information
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-1249-2015
- openFDA file date
- Nov 27, 2026
- Loaded by RegDataLab
- Nov 28, 2026
Need the full FDA record for Microport Orthopedics INC.? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.
Question to ask in Claude
Use RegDataLab to give me an FDA risk summary for Microport Orthopedics INC.: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.