Z-1278-2022 Class I Terminated

Recalled by Getinge Usa Sales Inc — Wayne, NJ

FDA device recall Z-1278-2022 was initiated by Getinge Usa Sales Inc on May 2, 2022 and is designated Class I. Reason for recall: Potential for crack to form on the on/off switch of the suction unit, which may result in an unreasonable risk of harm to the patient. In the worst case, the switch can break, preventing the suction… The recall status is terminated (terminated Aug 6, 2024). Affected quantity: 17 units US.

Recall Details

Product Type
Devices
Report Date
Jun 29, 2022
Initiation Date
May 2, 2022
Termination Date
Aug 6, 2024
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
17 units US

Product Description

Flow-c Anesthesia System : intended for use in administering anesthesia while controlling the entire ventilation of patient Product Code/REF Number: 6887700

Reason for Recall

Potential for crack to form on the on/off switch of the suction unit, which may result in an unreasonable risk of harm to the patient. In the worst case, the switch can break, preventing the suction unit from being activated.

Distribution Pattern

US Distribution in states of Colorado, Florida, New Jersey, Indiana, Missouri, and Virginia

Code Information

UDI-DI: 07325710009765 Serial Numbers: 4003, 4777, 4783, 5024, 5025, 5026, 5027, 5028, 5043, 5044, 5045, 5046, 5047, 5048, 5049, 5050, 5078. (Note: UDI was corrected July 7, 2022)

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1278-2022
openFDA file date
Oct 6, 2026
Loaded by RegDataLab
Oct 12, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it. FDA may have changed the recall since.