Z-1320-2018 Class II Terminated
FDA device recall Z-1320-2018 was initiated by Mindray DS USA, Inc. Dba Mindray North America on Jan 24, 2018 and is designated Class II. Reason for recall: Certain lots of M-30D diluent used with BC-3200 and BC-3600 Hematology Analyzers and M-53D diluent used with BC-5390 Hematology Analyzers may cause the system to produce an elevated platelet… The recall status is terminated (terminated Apr 19, 2019). Affected quantity: 4838.
Recall details
- Product Type
- Devices
- Report Date
- Apr 18, 2018
- Initiation Date
- Jan 24, 2018
- Termination Date
- Apr 19, 2019
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 4838
Product description
Hematology Diluent M-30D used with BC-3200 and BC-3600 Hematology Analyzer; Part Numbers: (1) 105-005405-00 (2 x 5.5L container), (2) 105-005406-00 (20L container) The M-30D Diluent is an azide-free, filtered isotonic solution for counting and sizing blood cells. It is for use with the Mindray BC-3200 and BC-3600 Hematology Analyzer.
Reason for recall
Certain lots of M-30D diluent used with BC-3200 and BC-3600 Hematology Analyzers and M-53D diluent used with BC-5390 Hematology Analyzers may cause the system to produce an elevated platelet background count.
Distribution pattern
US Distribution including Puerto Rico and to the states of : CA, MI, NC, TX, and WA. .
Code information
(1) 105-005405-00 (2 x 5.5L container) Lot Numbers: 2017060906, 2017041207, 2017042707, 2017041104, 2016122301, 2016112901, 2016060401, 2016052701, 2016052101, 2016050301; (2)105-005406-00 (20L container) Lot Numbers: 2017041104, 2016121901, 2016051801, 2016050301