Z-1321-2018 Class II Terminated
FDA device recall Z-1321-2018 was initiated by Mindray DS USA, Inc. Dba Mindray North America on Jan 24, 2018 and is designated Class II. Reason for recall: Certain lots of M-30D diluent used with BC-3200 and BC-3600 Hematology Analyzers and M-53D diluent used with BC-5390 Hematology Analyzer may cause the system to produce an elevated platelet background… The recall status is terminated (terminated Apr 19, 2019). Affected quantity: 496.
Recall details
- Product Type
- Devices
- Report Date
- Apr 18, 2018
- Initiation Date
- Jan 24, 2018
- Termination Date
- Apr 19, 2019
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 496
Product description
Hematology Diluent M-53D used with BC-5390 Hematology Analyzer; Part Number: 105-007867-00 (20L container) The M-53D diluent is an azide-free, filtered isotonic solution for counting and sizing blood cells. It is for use with the Mindray BC-5390 Hematology Analyzer.
Reason for recall
Certain lots of M-30D diluent used with BC-3200 and BC-3600 Hematology Analyzers and M-53D diluent used with BC-5390 Hematology Analyzer may cause the system to produce an elevated platelet background count.
Distribution pattern
US Distribution including Puerto Rico and to the states of : CA, MI, NC, TX, and WA. .
Code information
Lot Numbers: 2017060105 ¿ 2017041106 ¿ 2017010701 ¿ 2016121401 ¿ 2016112301 ¿ 2016111302