Z-1346-2013 Class II Terminated

Recalled by Roche Molecular Systems, Inc. — Branchburg, NJ

FDA device recall Z-1346-2013 was initiated by Roche Molecular Systems, Inc. on Apr 1, 2013 and is designated Class II. Reason for recall: In rare instances, channel shifted results have been generated with cobas¿ 4800 assays run on v1.1.1 of the cobas¿ 4800 system. Raw data review indicates that, in these cases, all signals are shifted… The recall status is terminated (terminated Aug 7, 2014). Affected quantity: 425 instruments.

Recall details

Product Type
Devices
Report Date
May 29, 2013
Initiation Date
Apr 1, 2013
Termination Date
Aug 7, 2014
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
425 instruments

Product description

cobas¿ 4800 system z480 instrument. Integrates fully automated total nucleic acid isolation directly from secondary sample tubes.

Reason for recall

In rare instances, channel shifted results have been generated with cobas¿ 4800 assays run on v1.1.1 of the cobas¿ 4800 system. Raw data review indicates that, in these cases, all signals are shifted by one channel: channel 1 results report as channel 2, channel 2 results report as channel 3, channel 3 results report as channel 4, and channel 4 results report as channel 1. In the reported

Distribution pattern

Nationwide Distribution.

Code information

Material Number 05200881001 - all lot/serial numbers.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1346-2013
openFDA file date
Nov 16, 2026
Loaded by RegDataLab
Nov 17, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-1346-2013. Report a data issue: hello@regdatalab.com.