Z-1346-2013 Class II Terminated
FDA device recall Z-1346-2013 was initiated by Roche Molecular Systems, Inc. on Apr 1, 2013 and is designated Class II. Reason for recall: In rare instances, channel shifted results have been generated with cobas¿ 4800 assays run on v1.1.1 of the cobas¿ 4800 system. Raw data review indicates that, in these cases, all signals are shifted… The recall status is terminated (terminated Aug 7, 2014). Affected quantity: 425 instruments.
Recall details
- Product Type
- Devices
- Report Date
- May 29, 2013
- Initiation Date
- Apr 1, 2013
- Termination Date
- Aug 7, 2014
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 425 instruments
Product description
cobas¿ 4800 system z480 instrument. Integrates fully automated total nucleic acid isolation directly from secondary sample tubes.
Reason for recall
In rare instances, channel shifted results have been generated with cobas¿ 4800 assays run on v1.1.1 of the cobas¿ 4800 system. Raw data review indicates that, in these cases, all signals are shifted by one channel: channel 1 results report as channel 2, channel 2 results report as channel 3, channel 3 results report as channel 4, and channel 4 results report as channel 1. In the reported
Distribution pattern
Nationwide Distribution.
Code information
Material Number 05200881001 - all lot/serial numbers.
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-1346-2013
- openFDA file date
- Nov 16, 2026
- Loaded by RegDataLab