Z-1370-2021 Class II Terminated

Recalled by Beckman Coulter Inc. — Brea, CA

Recall Details

Product Type
Devices
Report Date
April 14, 2021
Initiation Date
September 4, 2019
Termination Date
January 11, 2023
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
16 systems

Product Description

MicroScan WalkAway-40 plus Instrument

Reason for Recall

Instruments were manufactured with a shortened 30mL reagent draw straw which may result in reagents not dispensing and may lead to organism misidentification and/or Quality Control (QC) failure.

Distribution Pattern

U.S.: CO, FL, IL, IN, KY, ME, MI, MS, PA, Puerto Rico, TN, TX, UT, VA and WA O.U.S.: India, Indonesia, Italy, Japan, Philippines, Poland, Russia, South Korea, Taiwan and Yemen.

Code Information

Catalog Number: B1018-283 UDI Code: 1509959065866 Serial Numbers: 34010001 34010002 34010003 34010004 34010005 34010006 34010007 34010008 34010009 34010010 34010011 34010012 34010013 34010014 34010015 34010016