Z-1371-2021 Class II Terminated

Recalled by Beckman Coulter Inc. — Brea, CA

Recall Details

Product Type
Devices
Report Date
April 14, 2021
Initiation Date
September 4, 2019
Termination Date
January 11, 2023
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
12 systems

Product Description

MicroScan WalkAway-96 plus Instrument

Reason for Recall

Instruments were manufactured with a shortened 30mL reagent draw straw which may result in reagents not dispensing and may lead to organism misidentification and/or Quality Control (QC) failure.

Distribution Pattern

U.S.: CO, FL, IL, IN, KY, ME, MI, MS, PA, Puerto Rico, TN, TX, UT, VA and WA O.U.S.: India, Indonesia, Italy, Japan, Philippines, Poland, Russia, South Korea, Taiwan and Yemen.

Code Information

Catalog Number: B1018-284 UDI Code: 15099590658687 Serial Numbers: 39610001 39610002 39610003 39610004 39610005 39610006 39610007 39610008 39610009 39610010 39610011 39610012