Z-1373-2021 Class II Terminated

Recalled by Beckman Coulter Inc. — Brea, CA

FDA device recall Z-1373-2021 was initiated by Beckman Coulter Inc. on Sep 4, 2019 and is designated Class II. Reason for recall: Instruments were manufactured with a shortened 30mL reagent draw straw which may result in reagents not dispensing and may lead to organism misidentification and/or Quality Control (QC) failure. The recall status is terminated (terminated Jan 11, 2023). Affected quantity: 9 systems.

Recall details

Product Type
Devices
Report Date
Apr 14, 2021
Initiation Date
Sep 4, 2019
Termination Date
Jan 11, 2023
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
9 systems

Product description

MicroScan WalkAway-96 plus Reconditioned Instrument

Reason for recall

Instruments were manufactured with a shortened 30mL reagent draw straw which may result in reagents not dispensing and may lead to organism misidentification and/or Quality Control (QC) failure.

Distribution pattern

U.S.: CO, FL, IL, IN, KY, ME, MI, MS, PA, Puerto Rico, TN, TX, UT, VA and WA O.U.S.: India, Indonesia, Italy, Japan, Philippines, Poland, Russia, South Korea, Taiwan and Yemen.

Code information

Catalog Number: B1018-396R UDI Code: 15099590684129 Serial Numbers: 3967245 3967417 3967493 3967544 3967634 3967734 3967914 3928224 3968442

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1373-2021
openFDA file date
Nov 27, 2026
Loaded by RegDataLab
Nov 28, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Beckman Coulter Inc.? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Beckman Coulter Inc.: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Beckman Coulter Inc. →

Source: openFDA record for Z-1373-2021. Report a data issue: hello@regdatalab.com.