Z-1412-2015 Class II Terminated

Recalled by Natus Medical Incorporated — Seattle, WA

FDA device recall Z-1412-2015 was initiated by Natus Medical Incorporated on Mar 16, 2015 and is designated Class II. Reason for recall: neoBLUE blanket LED Phototherapy System is recalled because Natus has received feedback from some customers about early failure of the neoBLUE blanket system fiberoptic pads. These failures involve… The recall status is terminated (terminated Dec 20, 2016). Affected quantity: 2611 units (1911 units in the US and 700 units outside the US).

Recall details

Product Type
Devices
Report Date
Apr 22, 2015
Initiation Date
Mar 16, 2015
Termination Date
Dec 20, 2016
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
2611 units (1911 units in the US and 700 units outside the US)

Product description

Natus neoBLUE blanket LED Phototherapy system which includes the following items: 006224 neoBLUE blanket LED Phototherapy Systems with Large blanket/pad 006254 neoBLUE blanket LED Light Box 006895 neoBLUE blanket LED Phototherapy Systems with Small blanket/pad 007299 neoBLUE blanket LED Phototherapy Systems with Large blanket/pad EUR 007300 neoBLUE blanket LED Phototherapy Systems with Small blanket/pad EUR 007296 neoBLUE blanket LED Phototherapy Systems with Large blanket/pad UK 007298 neoBLUE blanket LED Phototherapy Systems with Small blanket/pad UK 007301 neoBLUE blanket LED Phototherapy Systems with Large blanket/pad AUS 007302 neoBLUE blanket LED Phototherapy Systems with Small blanket/pad AUS

Reason for recall

neoBLUE blanket LED Phototherapy System is recalled because Natus has received feedback from some customers about early failure of the neoBLUE blanket system fiberoptic pads. These failures involve discoloration and melting of the fiberoptic bundle at the connector that is inserted in the neoBLUE blanket light box.

Distribution pattern

distributed worldwide

Code information

None

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1412-2015
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Natus Medical Incorporated? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Natus Medical Incorporated: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Natus Medical Incorporated →

Source: openFDA record for Z-1412-2015. Report a data issue: hello@regdatalab.com.