Z-1503-2014 Class II Terminated

Recalled by Philips Medical Systems, Inc. — Andover, MA

FDA device recall Z-1503-2014 was initiated by Philips Medical Systems, Inc. on Mar 17, 2014 and is designated Class II. Reason for recall: The footboard of Patient support for stitching can be fixed in upper position. When the fix clamp (hook) is not correctly adjusted into its correct position, the footboard is not securely held. The recall status is terminated (terminated Sep 16, 2020). Affected quantity: 985 devices.

Recall details

Product Type
Devices
Report Date
Apr 30, 2014
Initiation Date
Mar 17, 2014
Termination Date
Sep 16, 2020
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
985 devices

Product description

Digital Diagnost, stationary radiographic system; Models: 712020/22/25/26/50/52/54/55/57/60/62/80/81/82/85

Reason for recall

The footboard of Patient support for stitching can be fixed in upper position. When the fix clamp (hook) is not correctly adjusted into its correct position, the footboard is not securely held.

Distribution pattern

Nationwide: US: AL, AR, AZ, CA, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, MA, MD, ME, MN, MO, MS, MT, NC, ND, NH, NJ, NM, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VT, WY / Worldwide including Canada & Mexico

Code information

Units distributed from November 2006 through March 2014.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1503-2014
openFDA file date
Nov 27, 2026
Loaded by RegDataLab
Nov 28, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Philips Medical Systems, Inc.? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Philips Medical Systems, Inc.: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Philips Medical Systems, Inc. →

Source: openFDA record for Z-1503-2014. Report a data issue: hello@regdatalab.com.