Z-1536-2017 Class II Terminated

Recalled by Medical Components, Inc Dba MedComp — Harleysville, PA

FDA device recall Z-1536-2017 was initiated by Medical Components, Inc Dba MedComp on Feb 23, 2017 and is designated Class II. Reason for recall: The 9.6F port kits were packaged with the incorrect valved peelable introducer. The label states the kit contains a 10F valved peelable introducer. The kit is packaged with a 9F valved peelable… The recall status is terminated (terminated Feb 2, 2018). Affected quantity: 71 units.

Recall details

Product Type
Devices
Report Date
Mar 29, 2017
Initiation Date
Feb 23, 2017
Termination Date
Feb 2, 2018
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
71 units

Product description

9.6F Plastic Dignity Mid-Sized CT Port w/Pre-Attached Silicone Catheter The CT Power Injectable Implantable Infusion Ports are indicated for patient therapies requiring repeated access to the vascular system. The port system can be used for infusion of medications, I.V. fluids, parenteral nutrition solutions, blood products, and for the withdrawal of blood samples. When used with a power injectable needle, the Power Injectable Implantable Infusion Port device is indicated for power injection of contrast media

Reason for recall

The 9.6F port kits were packaged with the incorrect valved peelable introducer. The label states the kit contains a 10F valved peelable introducer. The kit is packaged with a 9F valved peelable introducer. The port lumen will not fit through the introducer during the insertion procedure.

Distribution pattern

Worldwide Distribution to PR and Panama

Code information

MRCTI96801 Lot# MHWQ060 S2 , UDI# 884908031232

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1536-2017
openFDA file date
Nov 27, 2026
Loaded by RegDataLab
Nov 28, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Medical Components, Inc Dba MedComp? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Medical Components, Inc Dba MedComp: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Medical Components, Inc Dba MedComp →

Source: openFDA record for Z-1536-2017. Report a data issue: hello@regdatalab.com.