Z-1561-2020 Class II Terminated
FDA device recall Z-1561-2020 was initiated by Dexcom Inc on Feb 3, 2020 and is designated Class II. Reason for recall: The company identified potential interference from hydroxyurea. Patient use of the anti-neoplastic drug may falsely elevate glucose readings on the Continuous Glucose Monitoring Systems. The recall status is terminated (terminated Jan 20, 2022). Affected quantity: 85529 Systems.
Recall details
- Product Type
- Devices
- Report Date
- Apr 1, 2020
- Initiation Date
- Feb 3, 2020
- Termination Date
- Jan 20, 2022
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 85529 Systems
Product description
Reason for recall
Distribution pattern
World-wide Distribution
Code information
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-1561-2020
- openFDA file date
- Nov 19, 2026
- Loaded by RegDataLab
- Nov 20, 2026