Z-1562-2020 Class II Terminated

Recalled by Dexcom Inc — San Diego, CA

Recall Details

Product Type
Devices
Report Date
April 1, 2020
Initiation Date
February 3, 2020
Termination Date
January 20, 2022
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
275583

Product Description

Dexcom G6 Continuous Glucose Monitoring System The Dexcom G4 PLATINUM Continuous Glucose Monitoring System is a glucose monitoring device indicated for detecting trends and tracking patterns in persons (age 18 and older) with diabetes.

Reason for Recall

The company identified potential interference from hydroxyurea. Patient use of the anti-neoplastic drug may falsely elevate glucose readings on the Continuous Glucose Monitoring Systems.

Distribution Pattern

World-wide Distribution

Code Information

Device Listing Number: D319131 Model: STS-OR-003 Description: G6 Sensor Kit 3 Pack GUDID Registered GTIN: 50386270000250 Lot Number: All Model: STS-OE-003 Description: G6 Sensor Kit 3 Pack Retail GUDID Registered GTIN: 00386270000866 Lot Number: All Model: STS-OM-003 Description: G6 Sensor Kit 3 Pack Medicare GUDID Registered GTIN: 00386270001047 Lot Number: All Model: STS-OM-003 Description: G6 Sensor Kit 3 Pack Pro Q GUDID Registered GTIN: 70386270000254 Lot Number: All