Z-1577-2023 Class II Ongoing
Recall Details
- Product Type
- Devices
- Report Date
- May 17, 2023
- Initiation Date
- March 22, 2023
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 5 analyzers
Product Description
STA Compact Max fully automatic clinical analyzer, Reference Numbers 58602 and 58612
Reason for Recall
An internal investigation identified a bug in this firmware version, resulting in the following:Intermittent shortened coagulation times and an increased frequency of occurrence of a technical error reported by the analyzer (error 13).
Distribution Pattern
US Nationwide distribution in the states of IL, MA, MN, NE, TX, and WI.
Code Information
UDI-DI: 036074505898; Serial Numbers: 5061471 5061474 0000007405 CF70015885 CF70015886