Z-1598-2023 Class II Ongoing

Recalled by Beckman Coulter, Inc. — Miami, FL

FDA device recall Z-1598-2023 was initiated by Beckman Coulter, Inc. on Apr 5, 2023 and is designated Class II. Reason for recall: Beckman Coulter has become aware of an intermittent issue in which the optional flag "Previous Sample Had Sperm" was enabled but not displayed so that a carryover event could have been investigated. The recall status is ongoing. Affected quantity: 355 units.

Recall details

Product Type
Devices
Report Date
May 24, 2023
Initiation Date
Apr 5, 2023
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
355 units

Product description

DxU 840m Iris Urine Microscopy Analyzer, Catalog Number C76947, in-vitro diagnostic device, urine analysis

Reason for recall

Beckman Coulter has become aware of an intermittent issue in which the optional flag "Previous Sample Had Sperm" was enabled but not displayed so that a carryover event could have been investigated.

Distribution pattern

Worldwide Distribution

Code information

UDI/DI 15099590748654, All analyzers with Windows 10, Software Versions 8.5, 8.5.1 and 8.6

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1598-2023
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Beckman Coulter, Inc.? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Beckman Coulter, Inc.: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Beckman Coulter, Inc. →

Source: openFDA record for Z-1598-2023. Report a data issue: hello@regdatalab.com.