Z-1675-2015 Class II Terminated
FDA device recall Z-1675-2015 was initiated by Mindray DS USA, Inc. Dba Mindray North America on Mar 12, 2015 and is designated Class II. Reason for recall: Software anomaly in the Panorama System software version 8.9 that manifests when the Panorama is in use with Mindray Passport V Monitor. Arrhythmia alarms which were previously displayed on both the… The recall status is terminated (terminated Mar 28, 2017). Affected quantity: 4 units.
Recall details
- Product Type
- Devices
- Report Date
- Jun 3, 2015
- Initiation Date
- Mar 12, 2015
- Termination Date
- Mar 28, 2017
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 4 units
Product description
Panorama Patient Monitoring Network, Multi- Parameter Patient Monitor (with Arrhythmia Detection and Alarm. Part number 0998-00-0708-01. The Panorama Network includes the Panorama Telemetry System, which acquires and monitors physiological data for ambulating patients within a defined coverage area.
Reason for recall
Software anomaly in the Panorama System software version 8.9 that manifests when the Panorama is in use with Mindray Passport V Monitor. Arrhythmia alarms which were previously displayed on both the Passport V and Panorama will not be displayed on the Panorama following a restart of either system (due to communication loss, restart, etc.).
Distribution pattern
Worldwide Distribution - US Nationwide in the state of TX.
Code information
Part No: 0998-00-0708-01 Software version 8.9