Z-1675-2015 Class II Terminated

Recalled by Mindray DS USA, Inc. Dba Mindray North America — Mahwah, NJ

FDA device recall Z-1675-2015 was initiated by Mindray DS USA, Inc. Dba Mindray North America on Mar 12, 2015 and is designated Class II. Reason for recall: Software anomaly in the Panorama System software version 8.9 that manifests when the Panorama is in use with Mindray Passport V Monitor. Arrhythmia alarms which were previously displayed on both the… The recall status is terminated (terminated Mar 28, 2017). Affected quantity: 4 units.

Recall details

Product Type
Devices
Report Date
Jun 3, 2015
Initiation Date
Mar 12, 2015
Termination Date
Mar 28, 2017
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
4 units

Product description

Panorama Patient Monitoring Network, Multi- Parameter Patient Monitor (with Arrhythmia Detection and Alarm. Part number 0998-00-0708-01. The Panorama Network includes the Panorama Telemetry System, which acquires and monitors physiological data for ambulating patients within a defined coverage area.

Reason for recall

Software anomaly in the Panorama System software version 8.9 that manifests when the Panorama is in use with Mindray Passport V Monitor. Arrhythmia alarms which were previously displayed on both the Passport V and Panorama will not be displayed on the Panorama following a restart of either system (due to communication loss, restart, etc.).

Distribution pattern

Worldwide Distribution - US Nationwide in the state of TX.

Code information

Part No: 0998-00-0708-01 Software version 8.9

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1675-2015
openFDA file date
Nov 10, 2026
Loaded by RegDataLab
Nov 11, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-1675-2015. Report a data issue: hello@regdatalab.com.