Z-1707-2014 Class II Terminated

Recalled by Brainlab AG — Feldkirchen

FDA device recall Z-1707-2014 was initiated by Brainlab AG on Apr 15, 2014 and is designated Class II. Reason for recall: When using a specific plan update workflow in combination with a non-Brainlab treatment planning system (TPS), there is a possibility in ExacTrac 5.5.5 to generate an unintended blend of two initially… The recall status is terminated (terminated Apr 24, 2017). Affected quantity: 40 systems (worldwide).

Recall details

Product Type
Devices
Report Date
Jun 11, 2014
Initiation Date
Apr 15, 2014
Termination Date
Apr 24, 2017
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
40 systems (worldwide)

Product description

ExacTrac 5.5, an Image Processing System used for patient positioning for radiation therapy or radiosurgery.

Reason for recall

When using a specific plan update workflow in combination with a non-Brainlab treatment planning system (TPS), there is a possibility in ExacTrac 5.5.5 to generate an unintended blend of two initially separate plans of the same patient.

Distribution pattern

Worldwide Distribution - US Nationwide in the states of AZ, AR, CA, FL, KY, MA, NE, NY, TN, TX and the countries of Argentina, Australia, Brazil, Finland, India, Japan, Netherlands, Oman, Russia, and Taiwan

Code information

ExacTrac software version 5.5.5 is included in the following model / catalog numbers: 1) 20818B ET SOFTWARE 5.5.5; 2) 20820 ET SOFTWARE UPDATE 5.0 TO 5.5; 3) 49905D EXACTRAC COMPLETE SYSTEM UPG 3.X TO 5.5; 4) 49908C EXACTRAC SOFTWARE UPGRADE 3.X TO 5.5; 5) 49909A EXACTRAC SOFTWARE UPGRADE 4.X TO 5.5; 6) 49918 ET SOFTWARE UPDATE 5.5.X TO 5.5.5; 7) 49919 ET SOFTWARE UPDATE 5.0.X TO 5.5.5; 8) 49660A ET INFRARED SOFTWARE 5.5.5; 9) 49961 ET DATA PREP/REVIEW SYSTEM; 10) 49986 ET SYSTEM UPGRADE 3.X TO 5.5 NEW VAULT; 11) 49987 ET SYSTEM UPGRADE 4.X TO 5.5 NEW VAULT; 12) 49988 ET SYSTEM UPGRADE 4.X TO 5.5 SAME VAULT; 13) 49989 ET SYSTEM UPGRADE 5.0 TO 5.5 NEW VAULT; 14) 49990 ET SYSTEM UPGRADE 5.0 TO 5.5 SAME VAULT; 15) 49991 ET SYSTEM UPGRADE 5.5 TO 5.5 NEW VAULT

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1707-2014
openFDA file date
Nov 19, 2026
Loaded by RegDataLab
Nov 20, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Brainlab AG? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-1707-2014. Report a data issue: hello@regdatalab.com.