Z-1748-2022 Class II Ongoing

Recalled by FUJIFILM Healthcare Americas Corporation — Lexington, MA

Recall Details

Product Type
Devices
Report Date
September 21, 2022
Initiation Date
July 12, 2022
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
46 units

Product Description

SCENARIA View: Software Version V1.08, V1.09B, V1.09C, V1.09D, V1.0B, V1.0C

Reason for Recall

Error 00003050 or 00003052 may occur during scanning, the captured images will not be reconstructed and the Raw Data will not be displayed in the work list.result, may result in the need for additional scanning to capture the images properly for reconstruction, thereby exposing the patient to additional radiation

Distribution Pattern

US Nationwide Distribution

Code Information

UDI (01)04580292766694(11)190926(21)V5001(250)G560045919 (01)04580292766694(11)191002(21)V5002(250)G560045719 (01)04580292766694(11)191008(21)V5004(250)G560045519 (01)04580292766694(11)191107(21)V5005(250)G560046119 (01)04580292766694(11)191113(21)V5006(250)G560048819 (01)04580292766694(11)191218(21)V5007(250)G560052019 (01)04580292766694(11)191220(21)V5008(250)G560052219 (01)04580292766694(11)200124(21)V5009(250)G560052819 (01)04580292766694(11)200124(21)V5010(250)G560052919 (01)04580292766694(11)200224(21)V5011(250)G560053419 (01)04580292766694(11)200514(21)V5017(250)G560053219 (01)04580292766694(11)200904(21)V5018(250)G1Z0169319 (01)04580292766694(11)200911(21)V5019(250)G560055619 (01)04580292766694(11)200915(21)V5020(250)G1Z0169519 (01)04580292766694(11)200925(21)V5021(250)G1Z0170519 (01)04580292766694(11)201001(21)V5022(250)G1Z0170619 (01)04580292766694(11)201009(21)V5023(250)G560059219 (01)04580292766694(11)201014(21)V5024(250)G1Z0000620 (01)04580292766694(11)201020(21)V5026(250)G560077419 (01)04580292766694(11)201020(21)V5027(250)G1Z0001220 (01)04580292766694(11)201029(21)V5028(250)G1Z0001320 (01)04580292766694(11)210108(21)V5031(250)G560001120 (01)04580292766694(11)210112(21)V5032(250)G560001220 (01)04580292766694(11)210115(21)V5033(250)G560001320 (01)04580292766694(11)210208(21)V5034(250)G560002120 (01)04580292766694(11)210216(21)V5035(250)G560002320 (01)04580292766694(11)210217(21)V5036(250)G560002420 (01)04580292766694(11)210222(21)V5037(250)G560002720 (01)04580292766694(11)210226(21)V5038(250)G560001020 (01)04580292766694(11)210410(21)V5039(250)G560003120 (01)04580292766694(11)210414(21)V5040(250)G560003220 (01)04580292766694(11)210518(21)V5041(250)G560003620 (01)04573596200544(11)210713(21)V5042(250)G560007520 (01)04573596200544(11)210713(21)V5043(250)G560003920 (01)04573596200544(11)210729(21)V5044(250)G1Z0233220 (01)04573596200544(11)210729(21)V5045(250)G1Z0245920 (01)04573596200544(11)210902(21)V5046(250)G1Z0127821 (01)04573596200544(11)210903(21)V5047(250)G1Z0127521 (01)04573596200544(11)210909(21)V5048(250)G1Z0128621 (01)04573596200544(11)210922(21)V5049(250)G1Z0128721 (01)04573596200544(11)211004(21)V5050(250)G1Z0130121 (01)04573596200544(11)220302(21)V5051(250)G1Z0208921 (01)04573596200544(11)220304(21)V5052(250)G1Z0209021 (01)04573596200544(11)220307(21)V5053(250)G1Z0214121 (01)04573596200544(11)220330(21)V5054(250)G1Z0236821 Serial Numbers: G560045919 G560045719 G560045519 G560046119 G560048819 G560052019 G560052219 G560052819 G560052919 G560053419 G560053219 G1Z0169319 G560055619 G1Z0169519 G1Z0170519 G1Z0170619 G560059219 G1Z0000620 G560077419 G1Z0001220 G1Z0001320 G560001120 G560001220 G560001320 G560002120 G560002320 G560002420 G560002720 G560001020 G560003120 G560003220 G560003620 G560007520 G560003920 G1Z0233220 G1Z0245920 G1Z0127821 G1Z0127521 G1Z0128621 G1Z0128721 G1Z0130121 G1Z0208921 G1Z0209021 G1Z0214121 G1Z0236821