Z-1748-2022 Class II Ongoing
Recalled by FUJIFILM Healthcare Americas Corporation
— Lexington, MA
FDA device recall Z-1748-2022 was initiated by FUJIFILM Healthcare Americas Corporation on Jul 12, 2022 and is designated Class II. Reason for recall: Error 00003050 or 00003052 may occur during scanning, the captured images will not be reconstructed and the Raw Data will not be displayed in the work list.result, may result in the need for… The recall status is ongoing. Affected quantity: 46 units.
Recall details
- Product Type
- Devices
- Report Date
- Sep 21, 2022
- Initiation Date
- Jul 12, 2022
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 46 units
Product description
SCENARIA View: Software Version V1.08, V1.09B, V1.09C, V1.09D, V1.0B, V1.0C
Reason for recall
Error 00003050 or 00003052 may occur during scanning, the captured images will not be reconstructed and the Raw Data will not be displayed in the work list.result, may result in the need for additional scanning to capture the images properly for reconstruction, thereby exposing the patient to additional radiation
Distribution pattern
US Nationwide Distribution
Code information
UDI (01)04580292766694(11)190926(21)V5001(250)G560045919 (01)04580292766694(11)191002(21)V5002(250)G560045719 (01)04580292766694(11)191008(21)V5004(250)G560045519 (01)04580292766694(11)191107(21)V5005(250)G560046119 (01)04580292766694(11)191113(21)V5006(250)G560048819 (01)04580292766694(11)191218(21)V5007(250)G560052019 (01)04580292766694(11)191220(21)V5008(250)G560052219 (01)04580292766694(11)200124(21)V5009(250)G560052819 (01)04580292766694(11)200124(21)V5010(250)G560052919 (01)04580292766694(11)200224(21)V5011(250)G560053419 (01)04580292766694(11)200514(21)V5017(250)G560053219 (01)04580292766694(11)200904(21)V5018(250)G1Z0169319 (01)04580292766694(11)200911(21)V5019(250)G560055619 (01)04580292766694(11)200915(21)V5020(250)G1Z0169519 (01)04580292766694(11)200925(21)V5021(250)G1Z0170519 (01)04580292766694(11)201001(21)V5022(250)G1Z0170619 (01)04580292766694(11)201009(21)V5023(250)G560059219 (01)04580292766694(11)201014(21)V5024(250)G1Z0000620 (01)04580292766694(11)201020(21)V5026(250)G560077419 (01)04580292766694(11)201020(21)V5027(250)G1Z0001220 (01)04580292766694(11)201029(21)V5028(250)G1Z0001320 (01)04580292766694(11)210108(21)V5031(250)G560001120 (01)04580292766694(11)210112(21)V5032(250)G560001220 (01)04580292766694(11)210115(21)V5033(250)G560001320 (01)04580292766694(11)210208(21)V5034(250)G560002120 (01)04580292766694(11)210216(21)V5035(250)G560002320 (01)04580292766694(11)210217(21)V5036(250)G560002420 (01)04580292766694(11)210222(21)V5037(250)G560002720 (01)04580292766694(11)210226(21)V5038(250)G560001020 (01)04580292766694(11)210410(21)V5039(250)G560003120 (01)04580292766694(11)210414(21)V5040(250)G560003220 (01)04580292766694(11)210518(21)V5041(250)G560003620 (01)04573596200544(11)210713(21)V5042(250)G560007520 (01)04573596200544(11)210713(21)V5043(250)G560003920 (01)04573596200544(11)210729(21)V5044(250)G1Z0233220 (01)04573596200544(11)210729(21)V5045(250)G1Z0245920 (01)04573596200544(11)210902(21)V5046(250)G1Z0127821 (01)04573596200544(11)210903(21)V5047(250)G1Z0127521 (01)04573596200544(11)210909(21)V5048(250)G1Z0128621 (01)04573596200544(11)210922(21)V5049(250)G1Z0128721 (01)04573596200544(11)211004(21)V5050(250)G1Z0130121 (01)04573596200544(11)220302(21)V5051(250)G1Z0208921 (01)04573596200544(11)220304(21)V5052(250)G1Z0209021 (01)04573596200544(11)220307(21)V5053(250)G1Z0214121 (01)04573596200544(11)220330(21)V5054(250)G1Z0236821 Serial Numbers: G560045919 G560045719 G560045519 G560046119 G560048819 G560052019 G560052219 G560052819 G560052919 G560053419 G560053219 G1Z0169319 G560055619 G1Z0169519 G1Z0170519 G1Z0170619 G560059219 G1Z0000620 G560077419 G1Z0001220 G1Z0001320 G560001120 G560001220 G560001320 G560002120 G560002320 G560002420 G560002720 G560001020 G560003120 G560003220 G560003620 G560007520 G560003920 G1Z0233220 G1Z0245920 G1Z0127821 G1Z0127521 G1Z0128621 G1Z0128721 G1Z0130121 G1Z0208921 G1Z0209021 G1Z0214121 G1Z0236821
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-1748-2022
- openFDA file date
- Nov 19, 2026
- Loaded by RegDataLab
- Nov 20, 2026
These dates say when openFDA published the data file and when RegDataLab last loaded it.
Need the full FDA record for FUJIFILM Healthcare Americas Corporation? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.
Get a free API key
How the connector works