Z-1751-2020 Class II Terminated

Recalled by Beckman Coulter Inc. — Brea, CA

Recall Details

Product Type
Devices
Report Date
April 29, 2020
Initiation Date
September 4, 2019
Termination Date
April 28, 2021
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1610 units

Product Description

Synchron Systems Opiate 2000 ng Reagent (OP2 1 x 250); catalogue no. 475006 for use in conjunction with UniCel DxC 600/800 System(s) and SYNCHRON Systems Drugs of Abuse Testing (DAT) Urine Calibrators. The assay provides a rapid screening procedure for presence of the analyte in urine.

Reason for Recall

Cross-reactivity testing may not meet the claims indicated in the Opiate 2000 ng (OP2) Instructions For Use. The presence of certain compounds in urine may result in false positive or false negative patient test results.

Distribution Pattern

Distributed throughout the United States and to Puerto Rico. States include: CA, NC, OR, MI, GE, AL, SC, TX, NJ, IA, MS, IN, MA, NC, OH, AZ, LA, MA, NY, PA, KY, WV, MI, WA, CO, IL, VA, FL, KS, MD, ID, DE, NV, OK, AR, HI, WI, NY.

Code Information

Catalog Number 475006 ; UDI: 15099590222031; Lot Numbers: M904147 and older including M810039, M901087 and M904147) and 4 lots OP2 (M805122, M810039, and M901087. .