Z-1799-2020 Class II Terminated

Recalled by Beckman Coulter Inc. — Brea, CA

FDA device recall Z-1799-2020 was initiated by Beckman Coulter Inc. on Jun 12, 2019 and is designated Class II. Reason for recall: The firm has identified that cross-reactivity testing may not meet the claims indicated in the Barbiturates Reagent Chemistry Information Sheet (CIS). The presence of certain compounds in urine may… The recall status is terminated (terminated Dec 11, 2020). Affected quantity: 10,145 units.

Recall details

Product Type
Devices
Report Date
May 6, 2020
Initiation Date
Jun 12, 2019
Termination Date
Dec 11, 2020
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
10,145 units

Product description

SYNCHRON Systems Barbiturates Reagent (BARB 1 X 250); Catalog No. 475012 - Product Usage: BARB reagent, in conjunction with UniCel DxC 600/800 System(s) and SYNCHRON Systems Drugs of Abuse Testing (DAT) Urine Calibrators, is intended for the qualitative determination of barbiturates in human urine at a cutoff value of 200 ng/mL The BARB assay provides a rapid screening procedure for determining the presence of barbiturates (BARB) and its metabolites in urine.

Reason for recall

The firm has identified that cross-reactivity testing may not meet the claims indicated in the Barbiturates Reagent Chemistry Information Sheet (CIS). The presence of certain compounds in urine may result in false positive or false negative results.

Distribution pattern

Worldwide distribution - US Nationwide distribution and to multiple countries.

Code information

SYNCRHON Systems BARB 1 X 250); Catalog No. 475012; UDI: 15099590223007; Lot Numbers M903174, M902167, M801045, M811093, M809230, M807116, M806172, M805116, M804028, M803040, M802093 and older.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1799-2020
openFDA file date
Nov 27, 2026
Loaded by RegDataLab
Nov 28, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Beckman Coulter Inc.? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Beckman Coulter Inc.: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Beckman Coulter Inc. →

Source: openFDA record for Z-1799-2020. Report a data issue: hello@regdatalab.com.