Z-1831-2012 Class II Terminated

Recalled by Spacelabs Healthcare, Llc — Issaquah, WA

FDA device recall Z-1831-2012 was initiated by Spacelabs Healthcare, Llc on May 11, 2012 and is designated Class II. Reason for recall: The firm received reports of the Pathfinder LS software crashing after editing a batch of Holter Analyzer records. The software will stop with an error message requiring a restart. In addition, QTc… The recall status is terminated (terminated Oct 22, 2012). Affected quantity: 114.

Recall details

Product Type
Devices
Report Date
Jun 27, 2012
Initiation Date
May 11, 2012
Termination Date
Oct 22, 2012
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
114

Product description

The product is Pathfinder SL, version 1.6.0.3057 software used with the Spacelabs Healthcare Pathfinder SL Holter Analyzer, version 1.6.0. Product Usage: The Holter Analyzer analyzes information collected from ambulatory electrocardiogram recorders worn by patients and used to assist physicians in diagnosis and patient monitoring. In addition, incorrect QTc values are reported after moving the cursor. QTc values are timing events between labeled points of interest in the ECG signal that the physician can mark with the cursor. The software is used to edit a family of events and upon conclusion the software crashes with an error message.

Reason for recall

The firm received reports of the Pathfinder LS software crashing after editing a batch of Holter Analyzer records. The software will stop with an error message requiring a restart. In addition, QTc values will not update after moving the marker.

Distribution pattern

Worldwide Distribution - US (nationwide) including the states of Nevada and Vermont, and the countries of AUSTRALIA, AUSTRIA, CANADA, CYPRUS, GERMANY, ICELAND, ISRAEL, POLAND, ROMANIA, SOUTH AFRICA, SWITZERLAND, THAILAND, TURKEY and UNITED KINGDOM.

Code information

Version 1.6.0.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1831-2012
openFDA file date
Nov 27, 2026
Loaded by RegDataLab
Nov 28, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Spacelabs Healthcare, Llc? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Spacelabs Healthcare, Llc: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Spacelabs Healthcare, Llc →

Source: openFDA record for Z-1831-2012. Report a data issue: hello@regdatalab.com.