Z-1906-2025 Class II Ongoing
FDA device recall Z-1906-2025 was initiated by Dexcom, Inc. on May 5, 2025 and is designated Class II. Reason for recall: A bug in the 2.7.0 version of the G7 iOS App can cause the app to enter a state where it stops updating the estimated glucose value (EGV) and continues to show this stale EGV. If the user is unaware t… The recall status is ongoing. Affected quantity: 61,355 users.
Recall Details
- Product Type
- Devices
- Report Date
- June 11, 2025
- Initiation Date
- May 5, 2025
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 61,355 users
Product Description
Reason for Recall
Distribution Pattern
Worldwide Distribution: US (nationwide) and OUS (foreign) to countries of: United Arab Emirates (AE), Austria (AT), Australia (AU), Belgium (BE), Bahrain (BH), Canada (CA), Switzerland (CH), Czech Republic (CZ), Germany (DE), Denmark (DK), Spain (ES), Finland (FI), France (FR), United Kingdom (GB), Hong Kong (HK), Ireland (IE), Israel (IL), Iceland (IS), Italy (IT), Jordan (JO), Korea, Repulic (KR), Kuwait (KW), Lebanon (LB), Lithuania (LT), Luxemburg (LU), Malta (MT), Netherlands (NL), Norway (NO), New Zealand (NZ), Oman (OM), Poland (PL), Portugal (PT), Qatar (QA), Saudi Arabia (SA), Sweden (SE), Slovenia (SI), Slovakia (SK), Turkey (TR) and South Africa (ZA).