Z-1944-2020 Class II Terminated
Recall Details
- Product Type
- Devices
- Report Date
- May 20, 2020
- Initiation Date
- March 30, 2020
- Termination Date
- August 31, 2023
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 1557 units
Product Description
iChem¿VELOCITY Urine Chemistry System-
Reason for Recall
A defect in the SATA power adapter cable within the instruments computer has the potential risk for an electrical short that may lead to the following outcomes: charring and/or melting of the SATA adapter cable within the computers metal enclosure, smoke emanating from PC6, flames contained within PC6, or an inability to use the devices CD-ROM.
Distribution Pattern
US Distribution: Nationwide Foreign Distribution: Worldwide
Code Information
Catalog #700-7176-001 and #700-7177-001, All part numbers