Z-1945-2020 Class II Terminated

Recalled by Beckman Coulter Inc. — Brea, CA

Recall Details

Product Type
Devices
Report Date
May 20, 2020
Initiation Date
March 30, 2020
Termination Date
August 31, 2023
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
3576 units

Product Description

iQ200 Series Urine Microscopy Analyzer- All part numbers

Reason for Recall

A defect in the SATA power adapter cable within the instrument s computer has the potential risk for an electrical short that may lead to the following outcomes: charring and/or melting of the SATA adapter cable within the computer s metal enclosure, smoke emanating from PC6, flames contained within PC6, or an inability to use the device s CD-ROM.

Distribution Pattern

US Distribution: Nationwide Foreign Distribution: Worldwide

Code Information

Catalog Numbers: 00-3322, 700-3325, 700-3326, 700-3345, 700-3347, 700-3370, 700-3375, 700-3300RFB, B48999, C10683, C10684., All part numbers