Z-1995-2014 Class II Terminated

Recalled by Cenorin, LLC — Kent, WA

Recall Details

Product Type
Devices
Report Date
July 16, 2014
Initiation Date
January 24, 2013
Termination Date
February 10, 2015
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
25 HLD systems

Product Description

HLD System, Model 610, Medical Device Cleaning and High Level Disinfection Washer/Pasteurizer.

Reason for Recall

The temperature sensor/control system in the HLD Systems Model 610 may provide a false temperature reading. This caused the water to remain at ambient temperature throughout the pasteurization cycle rather than using hot water.

Distribution Pattern

Worldwide Distribution - US (nationwide): CA, DE, FL, MA, MI, NY, PA, TX, UT, WA, and WI and the countries of Uruguay and Taiwan.

Code Information

610601, 610603, 610703, 610704, 610705, 610707, 610708, 610802, 610803, 610804, 610901, 610903, 610904, 610905, 611001, 611002, 611004, 611101, 611102, 611103, 611201, 611202, 611203, 610902L, 611204.