Z-3033-2017 Class II Terminated

Recalled by Cenorin, LLC — Kent, WA

FDA device recall Z-3033-2017 was initiated by Cenorin, LLC on Jun 7, 2017 and is designated Class II. Reason for recall: Dryer unit not heating. Manufacturing issue causes the terminal block to overheat, char and/or burn. The recall status is terminated (terminated May 2, 2018). Affected quantity: 43 units.

Recall details

Product Type
Devices
Report Date
Aug 23, 2017
Initiation Date
Jun 7, 2017
Termination Date
May 2, 2018
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
43 units

Product description

ThermaSure(TM), a CENORIN(TM) product, SERIES 300 MEDICAL DEVICE DRYER: (a) Model 330; (b) Model 340 Forced air dryer for medical devices

Reason for recall

Dryer unit not heating. Manufacturing issue causes the terminal block to overheat, char and/or burn.

Distribution pattern

Worldwide Distribution - US (Nationwide) Internationally to Japan, Hong Kong, Brunei, and Israel

Code information

(a) Serial Numbers: 331517, 331518, 331523, 331524, 331525, 331526, 331527, 331528, 331529, 331530, 331601, 331608, 331609, 331612, 331615, 331616, 331621, 331622, 331623, 331624, 331625, 331626, 331627 (b) Serial numbers: 341506, 341507, 341516, 341520, 341521, 341522, 341602, 341603, 341604, 341605, 341606, 341607, 341610, 341611, 341613, 341614, 341617, 341618, 341619, 341620

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-3033-2017
openFDA file date
Nov 19, 2026
Loaded by RegDataLab
Nov 20, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Cenorin, LLC? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-3033-2017. Report a data issue: hello@regdatalab.com.