Z-2288-2018 Class II Terminated

Recalled by Cenorin, LLC — Kent, WA

FDA device recall Z-2288-2018 was initiated by Cenorin, LLC on Apr 9, 2018 and is designated Class II. Reason for recall: There is a supplier quality issue with the terminal blocks which may cause the terminal block to overheat, char and/or burn. The recall status is terminated (terminated Sep 13, 2019). Affected quantity: 14 units.

Recall details

Product Type
Devices
Report Date
Jul 4, 2018
Initiation Date
Apr 9, 2018
Termination Date
Sep 13, 2019
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
14 units

Product description

HLD Systems 600 Series Washer/Pasteurizer, UDI (01) 0085249007009 (11) 160929 (21) 61160; Model 610HT, 115V & Model 610HT, 230V Product Usage: Clean and thermally disinfect medical devices using full immersion pasteurization

Reason for recall

There is a supplier quality issue with the terminal blocks which may cause the terminal block to overheat, char and/or burn.

Distribution pattern

Worldwide Distribution - US Nationwide in the states of CA, FL, IN, NC, NE, UT & WV; and countries of Malaysia and Taiwan

Code information

Serial Numbers: 611503, 611504, 611505, 611506, 611601, 611602, 611603, 611604, 611605, 611606, 611607, 611608, 611609 & 611610

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2288-2018
openFDA file date
Nov 19, 2026
Loaded by RegDataLab
Nov 20, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Cenorin, LLC? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-2288-2018. Report a data issue: hello@regdatalab.com.