Z-2019-2012 Class II Terminated
Recalled by Spacelabs Healthcare, Llc
— Issaquah, WA
FDA device recall Z-2019-2012 was initiated by Spacelabs Healthcare, Llc on Jun 12, 2012 and is designated Class II. Reason for recall: There were reports that a certain combination of key presses will cause the module to disconnect from the Xprezzon bedside monitor, model 91393, and the patient parameters will drop from the display… The recall status is terminated (terminated Nov 28, 2012). Affected quantity: 21 units.
Recall details
- Product Type
- Devices
- Report Date
- Jul 25, 2012
- Initiation Date
- Jun 12, 2012
- Termination Date
- Nov 28, 2012
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 21 units
Product description
Spacelabs Medical Xprezzon Bedside Monitor, Model 91393, installations with Option C. It is a patient monitor with arrhythmia detection or alarms. The Spacelabs Healthcare Xprezzon Bedside Monitor is intended to alert the user to alarm conditions that are reported by Spacelabs Healthcare parameter modules and/or other devices via Flexport interfaces. The patient monitors are also capable of displaying alarm conditions on other monitors that are on the network through the Alarm Watch feature. The Spacelabs Healthcare Xprezzon Bedside Monitor may also function as a generic display or computer terminal. As a generic display or terminal, the patient monitor allows network based applications to open windows and display information the Xprezzon and other networked monitors. The Spacelabs Healthcare Xprezzon Bedside Monitor is designed to communicate with a variety of external devices such as displays, network devices, serial devices, user input devices, audio systems, and local/remote recorders. The Xprezzon Bedside Monitors is intended for use under the direct supervision of a licensed healthcare practitioner, or by personnel trained in proper use of the equipment in a hospital environment.
Reason for recall
There were reports that a certain combination of key presses will cause the module to disconnect from the Xprezzon bedside monitor, model 91393, and the patient parameters will drop from the display. The parameters return after power cycling the monitor.
Distribution pattern
Worldwide Distribution - USA including Florida, Iowa, Oklahoma, South Caroline and the countries of India, Germany and Panama.
Code information
040-1632-00; 1393-000013; 1393-000039; 1393-000040; 1393-000041; 1393-000341; 1393-000382; 1393-000383; 1393-000384; 1393-000385; 1393-000386; 1393-000387; 1393-000388; 1393-000808; 1393-000831; 1393-001374; 1393-001375; 1393-001453; 1393-001990; 1393-001991; and 1393-001992.
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-2019-2012
- openFDA file date
- Nov 27, 2026
- Loaded by RegDataLab
- Nov 28, 2026
These dates say when openFDA published the data file and when RegDataLab last loaded it.
Need the full FDA record for Spacelabs Healthcare, Llc? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.
Question to ask in Claude
Use RegDataLab to give me an FDA risk summary for Spacelabs Healthcare, Llc: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.
Ask Claude about Spacelabs Healthcare, Llc →
Use it in ChatGPT
Get a free API key
How the connector works