Z-2133-2012 Class II Terminated

Recalled by Abbott Point Of Care Inc. — Princeton, NJ

FDA device recall Z-2133-2012 was initiated by Abbott Point Of Care Inc. on Jan 1, 2011 and is designated Class II. Reason for recall: i-Stat EG7 cartridges may generate falsely depressed PCO2 results for samples above 26 mm hg when stored at room temperature for four weeks or longer. The recall status is terminated (terminated May 22, 2013). Affected quantity: 52,750 cartridges.

Recall details

Product Type
Devices
Report Date
Aug 8, 2012
Initiation Date
Jan 1, 2011
Termination Date
May 22, 2013
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
52,750 cartridges

Product description

I-STAT EG7+ Cartridge Abbott Point of Care, Princeton, NJ. The i-Stat EG7 cartridge is intended to be used for quantitatively measure sodium, potassium, ionized calcium, oxygen, carbon dioxide, partial pressure in arterial, venous, or capillary whole blood.

Reason for recall

i-Stat EG7 cartridges may generate falsely depressed PCO2 results for samples above 26 mm hg when stored at room temperature for four weeks or longer.

Distribution pattern

USA ( nationwide )

Code information

List number 03P76-25 Lot number S10246A