Z-2142-2019 Class II Terminated
FDA device recall Z-2142-2019 was initiated by Biomet 3I, LLC on Feb 26, 2019 and is designated Class II. Reason for recall: During manufacturing, the screw stop ledge was not made to specification and location within the abutment, resulting in the screw head seating deeper in the abutment with no positive stop. The recall status is terminated (terminated Jun 15, 2020). Affected quantity: 6 units.
Recall details
- Product Type
- Devices
- Report Date
- Aug 14, 2019
- Initiation Date
- Feb 26, 2019
- Termination Date
- Jun 15, 2020
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 6 units
Product description
Reason for recall
Distribution pattern
Nationwide and Puerto Rico, Canada, and Australia
Code information
Source
- Dataset