Z-2142-2019 Class II Terminated

Recalled by Biomet 3I, LLC — Palm Beach Gardens, FL

FDA device recall Z-2142-2019 was initiated by Biomet 3I, LLC on Feb 26, 2019 and is designated Class II. Reason for recall: During manufacturing, the screw stop ledge was not made to specification and location within the abutment, resulting in the screw head seating deeper in the abutment with no positive stop. The recall status is terminated (terminated Jun 15, 2020). Affected quantity: 6 units.

Recall details

Product Type
Devices
Report Date
Aug 14, 2019
Initiation Date
Feb 26, 2019
Termination Date
Jun 15, 2020
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
6 units

Product description

CERTAIN(R) BELLATEK(R) TITANIUM ABUTMENT 3.4MM, Item Number IIAT3 - Product Usage: BIOMET3i Restorative products are intended for use as an accessory to endosseous dental implants for placement in the maxilla and mandible

Reason for recall

During manufacturing, the screw stop ledge was not made to specification and location within the abutment, resulting in the screw head seating deeper in the abutment with no positive stop.

Distribution pattern

Nationwide and Puerto Rico, Canada, and Australia

Code information

Lot Numbers/UDI: 8441907-1/(01)00844868031284(10)8441907-1 8441907-2/(01)00844868031284(10)8441907-2 8441907-3/(01)00844868031284(10)8441907-3 8442367-1/(01)00844868031284(10)8442367-1 8439134A-3/(01)00844868031284(10)8439134A-3 8439134A-4/(01)00844868031284(10)8439134A-4

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2142-2019
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Biomet 3I, LLC? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Biomet 3I, LLC: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Biomet 3I, LLC →

Source: openFDA record for Z-2142-2019. Report a data issue: hello@regdatalab.com.