Z-2144-2019 Class II Terminated
Recalled by Biomet 3I, LLC
— Palm Beach Gardens, FL
FDA device recall Z-2144-2019 was initiated by Biomet 3I, LLC on Feb 26, 2019 and is designated Class II. Reason for recall: During manufacturing, the screw stop ledge was not made to specification and location within the abutment, resulting in the screw head seating deeper in the abutment with no positive stop. The recall status is terminated (terminated Jun 15, 2020). Affected quantity: 49 units.
Recall details
- Product Type
- Devices
- Report Date
- Aug 14, 2019
- Initiation Date
- Feb 26, 2019
- Termination Date
- Jun 15, 2020
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 49 units
Product description
CERTAIN(R) BELLATEK(R) TITANIUM ABUTMENT 3.4MM, Item Number ILDAT3 - Product Usage:BIOMET3i Restorative products are intended for use as an accessory to endosseous dental implants for placement in the maxilla and mandible
Reason for recall
During manufacturing, the screw stop ledge was not made to specification and location within the abutment, resulting in the screw head seating deeper in the abutment with no positive stop.
Distribution pattern
Nationwide and Puerto Rico, Canada, and Australia
Code information
Lot Numbers/UDI: 8442579-2/(01)00844868031390(10)8442579-2 8442701-1/(01)00844868031390(10)8442701-1 8442701-3/(01)00844868031390(10)8442701-3 8442692-1/(01)00844868031390(10)8442692-1 8442681-1/(01)00844868031390(10)8442681-1 8442663-1/(01)00844868031390(10)8442663-1 8442688-1/(01)00844868031390(10)8442688-1 8442837-1/(01)00844868031390(10)8442837-1 8442837-2/(01)00844868031390(10)8442837-2 8442657-1/(01)00844868031390(10)8442657-1 8442657-2/(01)00844868031390(10)8442657-2 8442674-1/(01)00844868031390(10)8442674-1 6867796-1/(01)00844868031390(10)6867796-1 8442805-1/(01)00844868031390(10)8442805-1 8442895-1/(01)00844868031390(10)8442895-1 8442811-1/(01)00844868031390(10)8442811-1 8442841-1/(01)00844868031390(10)8442841-1 8442720-1/(01)00844868031390(10)8442720-1 8442799-2/(01)00844868031390(10)8442799-2 8442765-1/(01)00844868031390(10)8442765-1 8442914-2/(01)00844868031390(10)8442914-2 8442944-1/(01)00844868031390(10)8442944-1 8443029-2/(01)00844868031390(10)8443029-2 8442940-2/(01)00844868031390(10)8442940-2 8443035-1/(01)00844868031390(10)8443035-1 8442948-1/(01)00844868031390(10)8442948-1 8443044-1/(01)00844868031390(10)8443044-1 8442965-1/(01)00844868031390(10)8442965-1 8442954-1/(01)00844868031390(10)8442954-1 8442949-1/(01)00844868031390(10)8442949-1 8442970-1/(01)00844868031390(10)8442970-1 8442930-1/(01)00844868031390(10)8442930-1 8443242-1/(01)00844868031390(10)8443242-1 8443098-1/(01)00844868031390(10)8443098-1 8442937-1/(01)00844868031390(10)8442937-1 8443046-1/(01)00844868031390(10)8443046-1 8442927-1/(01)00844868031390(10)8442927-1 8442936-1/(01)00844868031390(10)8442936-1 8442935-1/(01)00844868031390(10)8442935-1 8442997-1/(01)00844868031390(10)8442997-1 8443187-1/(01)00844868031390(10)8443187-1 8443018-1/(01)00844868031390(10)8443018-1 8443010-1/(01)00844868031390(10)8443010-1 8443230-1/(01)00844868031390(10)8443230-1 8443102-1/(01)00844868031390(10)8443102-1 8443231-1/(01)00844868031390(10)8443231-1 8443193-1/(01)00844868031390(10)8443193-1 8443248-1/(01)00844868031390(10)8443248-1 8443213-1/(01)00844868031390(10)8443213-1
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-2144-2019
- openFDA file date
- Nov 25, 2026
- Loaded by RegDataLab
- Nov 26, 2026
These dates say when openFDA published the data file and when RegDataLab last loaded it.
Need the full FDA record for Biomet 3I, LLC? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.
Question to ask in Claude
Use RegDataLab to give me an FDA risk summary for Biomet 3I, LLC: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.
Ask Claude about Biomet 3I, LLC →
Use it in ChatGPT
Get a free API key
How the connector works